40+ jaar expertise in laboratoriumapparatuur
Service & onderhoud door gecertificeerde specialisten

n-SAFE Cyto | Class II 3-Filter Safety Cabinet

Article no.: 311 2000 0600

The n-Safe Cyto is a Class II safety workstation with triple filtration for cytotoxic applications. All air is double HEPA H14 filtered and the system complies with EN 12469 and DIN 12980. The H14 cartridge pre-filters are compact and easy to change. Working height is available with adjustable or electric underframe.

Class II 3-filter safety workstation according to EN 12469 and DIN 12980
Double HEPA H14 filtration of all air, including pre-filters
H14 cartridge pre-filters with low pressure drop for easy maintenance
Adjustable air velocities within specified ranges
Compact design with extra legroom for seated working
📐Dimensions (L×W×H)60 × 77.8 × 207.8 cm
⚖️Weight145 kg
View all specifications →

Description

Safe Working with Cytotoxic Substances

Protection for Operator, Product and Environment

The n-Safe Cyto is a Class II safety workstation with triple filtration. The system is tested according to EN 12469 and DIN 12980. All air is double HEPA filtered. This supports safe working with cytotoxic compounds.

H14 Cartridge Pre-Filters with Low Pressure Drop

The n-Safe Cyto uses easily replaceable H14 cartridge pre-filters. These pre-filters have an exceptionally low pressure drop. This reduces service costs and supports a low noise level. The system is designed for low energy consumption.

Compact Design with Extra Legroom

The compact pre-filter construction in the understructure creates plenty of legroom. This enables a comfortable seated working posture. The controls and options align with the n-Safe Pro approach. The window opening and air velocities are adjustable within the specified ranges.

Key Features

  • Class II 3-filter cabinet according to EN 12469 and DIN 12980
  • Double HEPA filtration of all air (H14, EN1822)
  • H14 cartridge pre-filters, easy to replace and compact
  • Air velocities: downflow 0.28 m/s (adjustable 0.25–0.53) and inflow 0.45 m/s (adjustable 0.45–0.55)
  • LED lighting 0–2000 lux, dimmable
  • Front window opening 200 mm (adjustable 160–330 mm)
  • Stand height: adjustable 700–1050 mm or electric 750–1050 mm

Specifications

Property Specification
Type Class II microbiological safety workstation, 3-filter
Standards EN 12469, DIN 12980
Downflow air velocity 0.28 m/s (adjustable 0.25–0.53)
Inflow air velocity 0.45 m/s (adjustable 0.45–0.55)
Deviation < 8%
Filter technology HEPA H14 (EN1822), 99.999% at 0.3 µm (main, exhaust and pre-filter)
Ventilation principles Downflow and exhaust (model dependent, see variants)
Lighting LED 0–2000 lux, dimmable
Electrical data 220–240 V / 50–60 Hz or 110–120 V / 50–60 Hz
Front window opening 200 mm (adjustable 160–330 mm)
Stand height Adjustable 700–1050 mm; electric 750–1050 mm
Minimum clearance 800 mm
Note Brochure does not contain data for the 2100 mm model. This is explicitly marked below.

 

SKU Model Outer dimensions excl. stand (W x D x H) Weight Interior volume (approx.) Additional specifications
311 2000 0600 n-Safe Cyto 600 600 × 778 × 1843 mm 145 kg 200–236 L Interior: 540 × 530(625) × 700 mm. Downflow 510 m³/h. Exhaust 180 m³/h. Noise 53 dB(A). Power from 60 W.
311 2000 0900 n-Safe Cyto 900 1003 × 799 × 1843 mm 210 kg 328–394 L Interior: 900 × 520(625) × 700 mm. Downflow 510 m³/h. Exhaust 290 m³/h. Noise 54 dB(A). Power from 90 W.
311 2000 1200 n-Safe Cyto 1200 1303 × 799 × 1843 mm 240 kg 437–525 L Interior: 1200 × 520(625) × 700 mm. Downflow 680 m³/h. Exhaust 390 m³/h. Noise 55 dB(A). Power from 100 W.
311 2000 1500 n-Safe Cyto 1500 1603 × 799 × 1843 mm 270 kg 546–656 L Interior: 1500 × 520(625) × 700 mm. Downflow 840 m³/h. Exhaust 480 m³/h. Noise 56 dB(A). Power from 130 W.
311 2000 1800 n-Safe Cyto 1800 1907 × 799 × 1329 mm 300 kg 655–788 L Interior: 1800 × 520(625) × 700 mm. Downflow 1010 m³/h. Exhaust 580 m³/h. Noise 57 dB(A). Power from 140 W.
311 2000 2100 n-Safe Cyto 2100