n-SAFE B2 | Class II B2 Biological Safety Cabinet

Article no.: 311 4000 1200

The n-Safe B2 is a Class II B2 safety workstation with 100% exhaust without recirculation. The system requires connection to an external extraction system. The cabinet complies with NSF/ANSI 49 and EN 12469. Available in 1200, 1500 and 1800 mm working widths.

Class II B2 with 100% exhaust
No recirculating air in the workspace
Compliant with NSF/ANSI 49 and EN 12469
HEPA H14 main and exhaust filtration
Available in 1200, 1500 and 1800 mm working widths
📐Dimensions (L×W×H)130.3 × 79.9 × 213.5 cm
⚖️Weight235 kg
View all specifications →

Description

n-Safe B2 Class II B2 Safety Workstation

100% Exhaust Without Recirculation

The n-Safe B2 is a Class II B2 microbiological safety workstation according to NSF/ANSI 49. The cabinet provides protection for product, operator and environment. The system operates with 100% exhaust and has no recirculating air. This makes it suitable for applications where complete exhaust is required.

The B2 version requires connection to an external extraction system. The cabinet does not have a built-in exhaust fan. Exhaust air is passively discharged via a ducting system. The design complies with EN 12469 for product and operator protection.

The vertical downflow is 0.28 m/s and is adjustable between 0.25 and 0.53 m/s. The inflow is 0.45 m/s and is adjustable between 0.45 and 0.55 m/s. The deviation remains within ±8 percent. The system is equipped with HEPA H14 main and exhaust filters according to EN 1822.

Key Features

  • Class II B2 safety workstation according to NSF/ANSI 49
  • Protection compliant with EN 12469
  • 100% exhaust without recirculation
  • Requires external ducting, no built-in exhaust fan
  • HEPA H14 filtration according to EN 1822
  • Dimmable LED lighting 0–2000 lux
  • Front window opening 200 mm, adjustable 160–250 mm

Specifications

Property Specification
Type Class II B2 microbiological safety workstation
Standards NSF/ANSI 49, EN 12469
Downflow air velocity 0.28 m/s (adjustable 0.25–0.53)
Inflow air velocity 0.45 m/s (adjustable 0.45–0.55)
Deviation ±8%
Downflow capacity 680–1010 m³/h (model dependent)
Exhaust flow rate 1070–1590 m³/h (passive, ducting required)
Noise level 46–47 dB(A) according to EN 12469
Power From 100 W
Filter technology HEPA H14 EN 1822, 99.999% at 0.3 µm
Pre-filter G4
Lighting LED 0–2000 lux, dimmable
Mains voltage 220–240 V / 50–60 Hz or 110–120 V / 50–60 Hz
Front window opening 200 mm (adjustable 160–250 mm)
Working height fixed stand 800 or 900 mm (±25 mm)
Working height adjustable 700–1050 mm
Working height electric 750–1050 mm
Minimum clearance 800 mm

 

SKU Model Outer dimensions excl. stand (W x D x H) Inner dimensions (W x D x H) Weight Downflow / Exhaust (m³/h)
311 4000 1200 n-Safe B2 1200 1303 × 799 × 1405 mm 1200 × 520(625) × 700 mm 235 kg 680 / 1070
311 4000 1500 n-Safe B2 1500 1603 × 799 × 1405 mm 1500 × 520(625) × 700 mm 265 kg 840 / 1320
311 4000 1800 n-Safe B2 1800 1907 × 799 × 1405 mm 1800 × 520(625) × 700 mm 295 kg 1010 / 1590