n-SAFE | Class III Safety Cabinet

Article no.: 311 6200 1000

The n-Safe Class III is a fully enclosed microbiological safety workstation according to EN 12469. The cabinet is configured as a glove box with HEPA H14 filtration for supply and exhaust. Double exhaust filtration is optionally available. Suitable for applications with maximum containment requirements.

Fully enclosed Class III cabinet
HEPA H14 filtration for supply and exhaust
Optional double HEPA filtration on exhaust
Gas-tight valves and transfer hatches possible
Designed according to EN 12469
📐Dimensions (L×W×H)100.3 × 79.9 × 205 cm
⚖️Weight210 kg
View all specifications →

Description

n-Safe Class III Glove Box

Fully Enclosed System for Maximum Operator Protection

The n-Safe Class III is a fully enclosed microbiological safety workstation according to EN 12469. The system is designed as a glove box with fixed glove ports. The cabinet provides maximum operator and environmental protection. The workspace is completely gas-tight.

Both supply and exhaust air are HEPA H14 filtered according to EN 1822. Double HEPA H14 filtration on the exhaust is optionally available. The cabinet can be equipped with gas-tight valves for supply and exhaust. Access to technical components is possible via the front.

Transfer hatches are optionally available on both sides. This enables safe introduction or removal of materials. Combination with a Class II safety workstation is possible. Dimensions and configurations can be adjusted on request.

Key Features

  • Class III microbiological safety workstation according to EN 12469
  • Fully enclosed glove chamber with large glove ports
  • HEPA H14 filtration for supply and exhaust
  • Optional double HEPA filtration on exhaust
  • Gas-tight valves possible for supply and exhaust
  • Transfer hatches optionally on both sides
  • Access to technical components via front panel

Specifications

Property Specification
Type Class III microbiological safety workstation
Standard EN 12469
Construction Fully enclosed stainless steel workspace
Supply filtration HEPA H14 EN 1822, 99.999% at 0.3 µm
Exhaust filtration HEPA H14 EN 1822, 99.999% at 0.3 µm
Double exhaust HEPA Optional
Inflow without gloves > 0.7 m/s
Exhaust flow rate 460 m³/h
Noise level < 49 dB(A) according to EN 12469
Lighting LED 0–2000 lux, dimmable
Mains voltage 220–240 V / 50–60 Hz or 110–120 V / 50–60 Hz
Working height fixed stand 800 or 900 mm (±25 mm)
Working height adjustable 700–1050 mm
Working height electric 750–1050 mm
Minimum clearance 800 mm
Power From 90 W

 

SKU Model Outer dimensions excl. stand (W x D x H) Inner dimensions (W x D x H) Weight Notes
311 6200 1000 n-Safe Class III 1000 1003 × 798 × 1290 mm 1000 × 750 × 700 mm 210 kg Front door 950 × 400 mm
311 6200 1800 n-Safe Class III 1800