The n-Safe Cyto is a Class II safety workstation with triple filtration for cytotoxic applications. All air is double HEPA H14 filtered and the system complies with EN 12469 and DIN 12980. The H14 cartridge pre-filters are compact and easy to change. Working height is available with adjustable or electric underframe.
The n-Safe Cyto is a Class II safety workstation with triple filtration. The system is tested according to EN 12469 and DIN 12980. All air is double HEPA filtered. This supports safe working with cytotoxic compounds.
The n-Safe Cyto uses easily replaceable H14 cartridge pre-filters. These pre-filters have an exceptionally low pressure drop. This reduces service costs and supports a low noise level. The system is designed for low energy consumption.
The compact pre-filter construction in the understructure creates plenty of legroom. This enables a comfortable seated working posture. The controls and options align with the n-Safe Pro approach. The window opening and air velocities are adjustable within the specified ranges.
| Property | Specification |
|---|---|
| Type | Class II microbiological safety workstation, 3-filter |
| Standards | EN 12469, DIN 12980 |
| Downflow air velocity | 0.28 m/s (adjustable 0.25–0.53) |
| Inflow air velocity | 0.45 m/s (adjustable 0.45–0.55) |
| Deviation | < 8% |
| Filter technology | HEPA H14 (EN1822), 99.999% at 0.3 µm (main, exhaust and pre-filter) |
| Ventilation principles | Downflow and exhaust (model dependent, see variants) |
| Lighting | LED 0–2000 lux, dimmable |
| Electrical data | 220–240 V / 50–60 Hz or 110–120 V / 50–60 Hz |
| Front window opening | 200 mm (adjustable 160–330 mm) |
| Stand height | Adjustable 700–1050 mm; electric 750–1050 mm |
| Minimum clearance | 800 mm |
| Note | Brochure does not contain data for the 2100 mm model. This is explicitly marked below. |
| SKU | Model | Outer dimensions excl. stand (W x D x H) | Weight | Interior volume (approx.) | Additional specifications |
|---|---|---|---|---|---|
| 311 2000 0600 | n-Safe Cyto 600 | 600 × 778 × 1843 mm | 145 kg | 200–236 L | Interior: 540 × 530(625) × 700 mm. Downflow 510 m³/h. Exhaust 180 m³/h. Noise 53 dB(A). Power from 60 W. |
| 311 2000 0900 | n-Safe Cyto 900 | 1003 × 799 × 1843 mm | 210 kg | 328–394 L | Interior: 900 × 520(625) × 700 mm. Downflow 510 m³/h. Exhaust 290 m³/h. Noise 54 dB(A). Power from 90 W. |
| 311 2000 1200 | n-Safe Cyto 1200 | 1303 × 799 × 1843 mm | 240 kg | 437–525 L | Interior: 1200 × 520(625) × 700 mm. Downflow 680 m³/h. Exhaust 390 m³/h. Noise 55 dB(A). Power from 100 W. |
| 311 2000 1500 | n-Safe Cyto 1500 | 1603 × 799 × 1843 mm | 270 kg | 546–656 L | Interior: 1500 × 520(625) × 700 mm. Downflow 840 m³/h. Exhaust 480 m³/h. Noise 56 dB(A). Power from 130 W. |
| 311 2000 1800 | n-Safe Cyto 1800 | 1907 × 799 × 1329 mm | 300 kg | 655–788 L | Interior: 1800 × 520(625) × 700 mm. Downflow 1010 m³/h. Exhaust 580 m³/h. Noise 57 dB(A). Power from 140 W. |
| 311 2000 2100 | n-Safe Cyto 2100 |